TEX-TECH — FFP2 respiratory mask high filtration efficiency
White masks with ‘TEX-TECH FFP2 Premium’ labelling. The number 1040 appears next to the CE mark on the masks.
- Risk level
- Serious risk
- Tyyppi
- Health risk / other
- Luokka
- Protective equipment
- Vaara
- The particle/filter retention of the material is insufficient (measured values: as low as 83%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. Furthermore, the physical resistance of the filtering material is too high (measured values of exhalation resistance: up to 3,8 mbar). Consequently, it may prevent a sufficient quantity of air to be exhaled, causing difficulties to breathe.
- Origin
- Czechia
⚖️ Toimenpiteet viranomaisten määräämässä järjestyksessä
- ▸ Measures ordered by economic operators (to: Retailer) Recall of the product from end users
- ▸ (to: Other) Ban on the marketing of the product and any accompanying measures
8594176575094
01102020
What is this? A TEX-TECH brand FFP2 mask sold in Germany. It is meant to block small particles from the air you breathe.
What's happening? Tests show the mask does not filter well enough. It may let too many particles through. It also makes breathing out too hard.
Does this affect me? If you bought this mask in Germany, it may not protect you well.
What should I do? Stop using this mask. Ask your pharmacist or doctor which masks are safe to use.
🤖 Tämä selkeäsanainen yhteenveto on tuotettu viranomaisen ilmoituksesta tekoälyä käyttäen. Se on tarkoitettu yleisluontoiseen tiedottamiseen — ei lääkärinlausunto. Terveydestäsi päättäessäsi ota aina yhteyttä apteekkiin tai lääkäriin ja lue alla oleva virallinen lähde.
🔬 Lääketieteelliset / tekniset tiedot ammattilaisille
Risk level: Serious risk · Type: Health risk / other · Category: Protective equipment · Hazard: The particle/filter retention of the material is insufficient (measured values: as low as 83%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. Furthermore, the physical resistance of the filtering material is too high (measured values of exhalation resistance: up to 3,8 mbar). Consequently, it may prevent a sufficient quantity of air to be exhaled, causing difficulties to breathe. · Origin: Czechia
Kaikki tämän sivun tiedot perustuvat virallisen viraston tiedotteeseen. Käännämme ja tiivistämme ne; emme lisää tai muokkaa faktoja.