Isu — Disposable protective mask
Protective respiratory filtration half-face mask according to EN 149.
- Risk level
- Serious risk
- Typ
- Health risk / other
- Kategorie
- Protective equipment
- Origin
- People's Republic of China
Hazard: The product bears a CE marking but is not certified as protective equipment by a relevant body. Additionally, the particle/filter retention of the material is insufficient (measured value ≤ 88 %), the elastic cords of the mask easily detach and the mask does not properly adapt to the face leading to an insufficient total filtration capacity (measured mean value ≤ 39 %). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures.
⚖️ Von Behörden angeordnete Maßnahmen
- ▸ (to: Other) Import rejected at border
What is this? This is a disposable face mask sold under the brand Isu. It has a CE mark but is not properly tested for safety.
What's happening? Tests show the mask does not filter out enough particles. The straps can break, and it does not fit well, making it unsafe.
Does this affect me? If you bought this mask in Belgium, it may not protect you well from germs.
What should I do? Stop using it and ask your pharmacist or doctor for advice on safer masks.
🤖 Diese allgemeinverständliche Zusammenfassung wird automatisch aus der offiziellen Mitteilung der Behörde mithilfe von KI erstellt. Sie dient nur zur allgemeinen Information – nicht als medizinischer Rat. Treffen Sie gesundheitsbezogene Entscheidungen immer in Absprache mit einer Apothekerin oder einem Arzt und lesen Sie die verlinkte offizielle Quelle unten.
🔬 Medizinische / technische Details Für Fachkräfte
Risk level: Serious risk · Type: Health risk / other · Category: Protective equipment · Hazard: The product bears a CE marking but is not certified as protective equipment by a relevant body. Additionally, the particle/filter retention of the material is insufficient (measured value ≤ 88 %), the elastic cords of the mask easily detach and the mask does not properly adapt to the face leading to an insufficient total filtration capacity (measured mean value ≤ 39 %). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. · Origin: People's Republic of China
Alle Informationen auf dieser Seite stammen aus der offiziellen Agenturmitteilung. Wir übersetzen und fassen sie zusammen; wir ergänzen oder ändern keine Fakten.