TEX-TECH — FFP2 respiratory mask high filtration efficiency
White masks with ‘TEX-TECH FFP2 Premium’ labelling. The number 1040 appears next to the CE mark on the masks.
- Risk level
- Serious risk
- Tipo
- Health risk / other
- Categoria
- Protective equipment
- Rischio
- The particle/filter retention of the material is insufficient (measured values: as low as 83%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. Furthermore, the physical resistance of the filtering material is too high (measured values of exhalation resistance: up to 3,8 mbar). Consequently, it may prevent a sufficient quantity of air to be exhaled, causing difficulties to breathe.
- Origin
- Czechia
⚖️ Azioni ordinate dalle autorità
- ▸ Measures ordered by economic operators (to: Retailer) Recall of the product from end users
- ▸ (to: Other) Ban on the marketing of the product and any accompanying measures
8594176575094
01102020
What is this? A TEX-TECH brand FFP2 mask sold in Germany. It is meant to block small particles from the air you breathe.
What's happening? Tests show the mask does not filter well enough. It may let too many particles through. It also makes breathing out too hard.
Does this affect me? If you bought this mask in Germany, it may not protect you well.
What should I do? Stop using this mask. Ask your pharmacist or doctor which masks are safe to use.
🤖 Questo riassunto in linguaggio semplice viene generato automaticamente dalla notifica ufficiale dell'agenzia tramite IA. È solo a scopo informativo generale — non è un consiglio medico. Per decisioni relative alla tua salute, consulta sempre un farmacista o un medico e leggi la fonte ufficiale collegata di seguito.
🔬 Dettagli medici/tecnici Per professionisti
Risk level: Serious risk · Type: Health risk / other · Category: Protective equipment · Hazard: The particle/filter retention of the material is insufficient (measured values: as low as 83%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. Furthermore, the physical resistance of the filtering material is too high (measured values of exhalation resistance: up to 3,8 mbar). Consequently, it may prevent a sufficient quantity of air to be exhaled, causing difficulties to breathe. · Origin: Czechia
Tutte le informazioni su questa pagina provengono dal comunicato ufficiale dell'agenzia. Le traduciamo e le riassumiamo; non aggiungiamo né modifichiamo i fatti.